11.4 Replacement of fuses

rd
A. Switch off the Chiropro 3
B. Disconnect the mains cable.
CAUTION
The power cable must be disconnected at least 10 seconds before
opening the fusebox.
FIG. 2
C. Remove the fuse box (1) with a flat screwdriver.
FIG. 3
D. Replace the fuses (2) by the new ones and put the fuse box
back (1) in place.
CAUTION
Only use fuses T4.0AH 250 VAC REF 1307312-010.
1
1
Gen unit.
2
FIG. 3
NOTES
1
Bien-Air Dental SA recommends the user to have its dynamic
instruments regularly checked or inspected.
2
The knob is hold magnetically. There is no need to preserve its
angular position when removing it or putting it back in place.