If points A2 and A3 are more than 0.2 mm / m apart on either the X, Y or Z axis,
!
the device is in need of adjustment. Contact your authorised dealer or
else the UMAREX-LASERLINER Service Department or follow the calibration
instructions in the manual.

Calibration procedure

6. Once you have completed steps 1 to 4 and identified dots A2 (reference) and A3
(deviation), use the corresponding screws to level the laser so that dot A3 covers
dot A2. During this process, the air bubble in the level will move out of its central
position and stray to the side.
7. Remove the protective caps from the ends of the levels to
expose the calibration screws. Turn the screws to calibrate
the level (using the SW 2.5 socket wrench included in the
scope of supply). Calibration is complete when the air bubble
returns to the centre of the level.
8. Repeat for the remaining axes.

Technical data

Accuracy
Laser class
Laser wavelength
Beam splitter
Reference beam
Horizontal plano
Speed of rotation
Battery life
Continious operation
Working- / Storage temperature
Weight
(Subject to technical alterations. 01.17)
± 0.2 mm / m
2 / < 1 mW
635 nm
20"
vertical
inclination can be adjusted manually
0 ... 120 rpm variable,
550 rpm for receiver
40 h (4 x 1.5 V of type AA Mignon alkaline)
connection socket for external supply
0°C to 50°C / -10°C to 70°C
0.8 kg
BeamControl-Master
Calibration screws