Install the device in a location and position that gives easy access to the disconnect device and its operation.
Supply power to the monitor with conduit. Do not use a power cord to supply power. Make sure that a
circuit breaker with sufficient current capacity is installed in the power line. The circuit breaker size is
based on the wire gauge used for installation.
• Install a local disconnect for the monitor within 3 m (10 ft) of the monitor. Put a label on the
disconnect that identifies it as the main disconnect device for the monitor.
• Make sure that the power and safety ground service drops for the monitor are 18–12 AWG and the
wire insulation is rated for 300 VAC or higher and 60 °C (140 °F) or higher.
• Connect equipment in accordance with local, state or national electrical codes.
• Connect the conduit through a conduit hub that holds the conduit securely and seals the enclosure
when tightened, and use the conduit hubs to keep the enclosure environmental rating.
• If metal conduit is used, make sure that the conduit hub is tightened so that the conduit hub
connects the metal conduit to safely ground.
Connect power
1. Turn the door latch with a wide flat-blade screwdriver or a coin and pliers. Open the door.
2. Install a conduit hub in the conduit opening. Refer to
conduit opening.
3. Put conduit (16–18 AWG) through the conduit hub.
4. Connect the conduit wires (16–18 AWG) to the J1 connector and the ground stud. Refer to
Figure 3
and
Table
5. Tighten the conduit hub.
6. Close the door.
7. Turn the door latch with a wide flat-blade screwdriver or a coin and pliers.
Figure 3 Connect power wiring
1 Hot wire
2 Neutral wire
3 Ground wire
8 English