5

Start-up

5.1

Transmitter settings

Activation of the WaveInjector mode
SYSTEM settings\Measuring
• Select Special Funct.\SYSTEM settings\Measuring.
For the selection of the menu items see operating instruction FLUXUS.
SYSTEM settings\Measuring\WaveInjector
• Press ENTER until WaveInjector is displayed.
• Select on to activate the WaveInjector mode.
• Press ENTER.
This display will only be indicated if the transmitter is equipped with the WaveInjector option.
Parameter input
• Select the program branch Parameter in the main menu.
• Press ENTER.
Parameter\for Channel A
• Select the measuring channel to which the transducers are connected.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Enter the pipe and fluid parameters, see operating instruction FLUXUS.
• Press ENTER after each input.
Notice!
The entered pipe parameters have to correspond to the parameters specified in the order.
Parameter\WaveInjector\WI-400
• Scroll through the scroll list and select the coupling plate of the WaveInjector, see Tab. 5.1.
• Press ENTER.
Parameter\WaveInjector\With Sensorbox?
• Select yes if the WaveInjector is to be used with the transducer box.
• Press ENTER.
92
Supplement to operating instruction - ENGLISH
2018-01-19, SUFLUXUS_WIV4-3EN