Consult the charger's individual instructions for further operational details,
additional warnings, and safety data.
6.2 BATTERY REPLACEMENT
Unscrew the headpiece that houses the LED. (Do not remove the switch; the battery
cannot fit through the switch end.) Remove the discharged battery and insert a fully
charged battery. Observe correct polarity. The positive (+) end of the battery must point
toward the headpiece. The positive end of the battery has a protruding nipple; the
negative (-) end is flat.
7.
CLEANING / STERILIZATION
7.1 CLEANING
The LED Rover IV+ is a high quality optical device. Care for it appropriately.
A soft brush with mild detergent may be used to clean the device. The lens of the LED
inside the optic port may be wiped with a swab. Avoid harsh or abrasive material that
could scratch or in any way damage the optical surface.
7.2
STERILIZATION
The LED Rover IV+ is provided non-sterile. The device should be cleaned and
disinfected or sterilized prior to use, according to your institution's requirements.
Compatible Sterilization Method: STERRAD™ in STERRAD sterilizer. Double
wrap with sterilization wrap. The full cycle of the Sterrad sterilizer comprises two
identical exposure phases (6 minute injection, 2 minute diffusion and 2 minute plasma)
preceded by a 10-minute pre-exposure plasma phase.
Follow the instructions provided with the STERRAD Sterilization System for
sterilization processing via this equipment and method.
NOTE: The LED Rover IV+ has a black anodized exterior finish which is affected
by Sterrad sterilization. The black anodized finish gradually bleaches; however,
this does not impact the functionality of the unit.
CAUTION
Do not reprocess the LED Rover IV+ using steam sterilization or autoclave. Use
of these processes will result in damage to the instrument, and void its warranty.
Process per your institution's validated procedures and process parameters.
7.3
HIGH LEVEL DISINFECTION (IMMERSION)
The LED Rover IV+ may be immersed in high level disinfectant solution no deeper than
six (6) inches (15cm) for as long as one (1) hour. Process according to the disinfectant
manufacturer's instructions or your institution's established protocol. We make no
claims other than the Rover IV+ immersibility as stated.
LIT-226 Sunoptic Technologies
Rev. B
®
Page 6 of 35