Page 3
1
Oload F
2
Oload 260V
3
Oload 280V
4
Oload MaxV
5
Oload T>2
6
Oload T>4
7
Oload T>8
8
Oload T>12
9
Oload T>16
10
Oload Temp
11
Angle LR
12
Time Sec
Chart 1 —
LINIS code chart as shown on LCD in siderails.
Project
SERVICE MANUAL
Date
08/2014
Version
08
Force overloads number
Overstrain number > 260V
Overstrain number > 280V
Max. overstrain value
Duty cycle exceeded
Duty cycle exceeded
Duty cycle exceeded
Duty cycle exceeded
Duty cycle exceeded
Heat overload value
(see 7.3)
Calfrest angle
RTC seconds
1
Scale C1
2
Scale C2
3
Scale C3
4
Scale C4
5
Scale CG
6
Scale CCal
7
Scale NoCal
8
Scale Graph
9
Scale T1
10
Scale T2
11
Scale T3
12
Scale T4
Tensometer constant 1
Tensometer constant 2
Tensometer constant 3
Tensometer constant 4
Gravitational constant
Calibration constant
Number of calibrations
Scale value saving
Tensometer value 1
Tensometer value 2
Tensometer value 3
Tensometer value 4
88/91