CHECKPOINT (P/N 9094) SPECIFICATIONS:
Stimulus
16Hz (+/- 1Hz)
Frequency
Stimulus
User Selectable:
Amplitude
Stimulus
Continuously adjustable from 0µs to 200µs (maximum
Pulse
setting is +/-20µs)
Duration
Stimulus
Biphasic stimulus waveform with controlled current
Waveform
during the cathodic (leading) phase and no net DC
current.
Specified operation is ensured into any patient circuit
load resistance up to 1,500Ω.
Operating
Approximately seven (7) hours of operation
Life
Power
Two size N alkaline cells (permanently sealed within
Source
device)
Operating
+16°C to +26°C (61°F to79°F)
Temperature
Relative
30% to 60%RH
Humidity
Atmospheric
88kPA to 106kPa
Pressure
Storage
-20°C to +55°C
Temperature
Product has no user serviceable or repairable parts.
Sterilized with Ethylene Oxide.
Protection against water ingress: Ordinary Equipment (IPX0).
Do NOT resterilize.
Rx Only
0.5mA (+/- 0.05mA)
2mA (+/- 0.2mA)
20mA (+/- 2mA)
Cathodic
Phase
Stimulus
Amplitude
(Current)
Pulse Duration
6 of 36
Anodic Phase
(Charge Recovery)
Interphase Delay
EN
4260-MAN-000-V