Disinfection:
Preparing the Resusci Manikin Face for reuse
Laerdal High Level Manikin Disinfectant
Note: Wear gloves* during cleaning.
1. Remove face and detach
the coupling from inside the face.
2. Clean and disinfect faces
and couplings as follows:
10 min.
5 litres
To remove the airway
1
Having removed face, remove neck
skin from chin.
5
Connect expiration and inspiration
tubes.
3
Push expiration tube into retainer in
manikin's side until end
protrudes about 2-3 cm (1").
Caution
Make sure that the SkillReporter/manikin has been turned off /
the Skillguide is unplugged from manikin:
- when airway is changed, or
- when internal service is carried out.
A f t e r c l a s s
Airway is designed for disposal after each class.
2
Pull to remove jaw.
6
Slide tubing through slot in lung
plate, to detach airway assembly.
Discard used airway.
4
Push inspiration tube through neck
and click airway connector into its
holder.
a) Use a 50g sachet of Laerdal High Level Manikin
Disinfectant (Virkon) which is enough for 5 litres
of water (1% solution).
b) Immerse all components into the solution.
c) Scrub stubborn stains where necessary and leave
for 10 minutes.
d) Rinse with fresh water.
3. Allow parts to dry.
4. Reassemble face coupling and face skin.
5. Remount face on manikin after reinstallation of airway
or put in a clean bag for future use.
* Vinyl gloves are recommended as latex gloves
may stain face mask.
3
Detach airway connector and pull
inspiration tube through the neck.
To install the airway
1
Slide tubing through slot in lung
plate, with lung on top.
5
Install jaw by sliding its forked
portions over steering pins
in the inner head.
To reinstall chest cover:
Make sure the switch cover for the
hand position impulse unit (a) is on top
of the protruding end of the slider for
the ventilation-compression impulse unit
(b). Button the chest cover onto torso
11
Resusci Anne Modular System
4
Pull expiration tube from its retainer
in the manikin's side.
2
Push on lung connector until fully
seated in lung plate.
Make sure the ventilation slider can
move freely.
6
Fold neck skin over jaw, and install
face.
a
b