Electric shock risk.
Operating field
Undesired effects of overlapping light fields
Possibility of tissue dehydration
and damage.
MU_023
Rev.2
MAINTENANCE MANUAL
1 GENERAL SAFETY INFORMATION
This manual is an integral part of the Product as indicated by
European Directive 93/42/EEC and subsequent amendments and
supplements. Read and keep this Operation and Maintenance
Manual close to the Product.
TECNO-GAZ disclaims all liability for any injury to persons or
damage to property caused by the USE or MAINTENANCE of the
Product by persons who are not OPERATORS or TECHNICAL SERVICE
PERSONNEL.
The Product is an EM electro-medical equipment and therefore falls
within the field of application of the EN/IEC 62353 standard.
To avoid any risk of electric shocks, the Product must only be
connected to mains supplies with earth protection.
2 Importance of personal safety
2.1 Intended use
SECONDARY SURGICAL LAMP (TREATMENT LAMP)
The Product is a medical device designed for use in operating
theatres within the PATIENT AREA, with short-term duration, active,
non invasive, designed to locally light up the patient's body for
treatments and diagnosis which can be interrupted without
DANGER for the PATIENT in case of a power outage.
A combination of two or more surgical lamps used in the operating
theatre and required for treatment and diagnosis makes up a
SURGICAL LAMP SYSTEM.
The Product correctly lights up the operating range from a distance
of about 70 – 140 cm from the patient area.
In the event of overlapping lamps, a temperature increase would
ensue in the patient area with consequent risk of dehydration and
tissue damage.
In case of a reduction in blood flow with start of tissue dehydration,
reduce light intensity.
28/10/2019
OPERATION AND
EN
Page 4 of 32