General data
Colour
Directive
Standards
Classification of Medical Device
does not vary by more than 20% during use; the colour temperature and the colour rendering
index are stable and are within the range 3000K-6700K and 85-100, respectively; E
Essential
performance
Limitation of energy in the operating field (UV-irradiance for wavelengths below 400 nm does not
exceed 10 W/m
IP degree of protection
Operating conditions
Handpiece steam sterilization
Mains power voltage insulation means
Dimensions
Diameter of lamp body [cm]
Light emission surface [cm
Weight of single ceiling, double ceiling, mobile,
mobile battery, wall surgical light [kg]
Markings
All technical light measurements are to be deemed with a tolerance of ±6% for metrological and manufacturing
reasons
MU_023
Rev.2
Distribution of minimum and adequate lighting (luminous flux emitted by the ME equipment
2
and the total irradiance Ee in the lighted area does not exceed 1000 W/m
distance of 1000 mm; E
2
]
28/10/2019
OPERATION AND
MAINTENANCE MANUAL
93/42/CEE and 2007/47/EC
be ≥ 40,000 lux and ≤ 160,000 lux).
value shall be ≥ 40,000 lux and ≤ 160,000 lux; E
c
121°C at 1.3bar from 25 to 30 minutes.
134°C at 2.3bar for 4 minutes.
Outside the product (main switch) for ceiling versions
Main switch for mobile and wall versions
In conformity with Directive 93/42/EEC (and 2007/47/EC)
RAL 9003
IEC 60601-2-41
Class I
/E
e
IP20
Continuous operation
40
305
13, 20, 21, 24, 12
EN
value shall
c
2
at a
≤ 6 mV/m
2
lx).
c
Page 24 of 32