Before initial use the product should be kept in its original packaging. GCE
recommends use of the original packaging (including internal sealing bag
and caps) if the product is withdraw from operation (for transport, storage).
Statutory laws, rules and regulations for medical gases, accident prevention
and environmental protection must be observed.
OPERATING CONDITIONS
EN
In case of storage at a temperature below -20 °C, d o not operate the re-
gulator until it has been allowed to increase its temperature to a minimum
of -20 °C.
At regulators intended for use with mixture of medical gases O
lower operation temperature limit +5°C. During normal use the flow outlet
and pressure outlet will sometimes have a frosty appearance. This is a
normal physical reaction in the valve, due to that the gas is going from
high pressure to low pressure (Joule Thompson effect). Ensure that all
equipment that is connected to the valve by at least a 2 metre hose.
4. PERSONNEL INSTRUCTIONS AND TRAINING
According to Medical Devices Directive 93/42/EEC the product provider
must ensure that all personnel handling the product are provided with the
operating instructions & performance data and are fully trained to carry out
that operation. Trainees need to be supervised by an experienced person.
Do not use the device without being trained. Training can be only done by
a person with an appropriate education, experience and knowledge that
has been also trained by the manufacturer.
For further information about training programs, please contact GCE.

5. PRODUCT DESCRIPTION

FIG. 1: Typical configuration of MediSelect® II regulator
FIG. 2: Configuration of MediReg® II regulator with flowmeter
The regulator acts as a pressure-reducer, gas from the cylinder valve
passes through the pressure regulator to the user outlets.
A - Inlet stem
Regulator is fitted to the medical cylinder valve by mean of an inlet stem.
The stem can be bull nose type (male thread), nut type (female thread) or
pin index type. The inlet stem includes a filter.
-20/+60 °C
10/100%
600/1200 mbar
STORAGE AND TRANSPORT
CONDITIONS
-30/+60 °C
10/100%
600/1200 mbar
4/195