Printing measurement results

Every device is equipped with Bluetooth, and can be configured accordingly if
necessary. To print the measurement results, the device can be connected to the
Dräger Mobile Printer Bluetooth
®
Bluetooth
.
Observe the Dräger Mobile Printer Bluetooth
®
Bluetooth
5

Troubleshooting

An error message is displayed in the event of instrument faults. The number that
appears below the message is used for service functions. If the fault remains
after switching the instrument on and off several times, contact DrägerService or
your dealer.
Fault/message
The instrument cannot
be switched on.
Volume too low /
sample invalid.
Instrument outside
temperature range
↑ ↑ ↑
(over range)
Volume too small
®
must not be selected in aircraft.
Cause
Battery drained.
Test subject is blowing
intermittently or not hard
enough.
Instrument is too hot or too
cold.
Measuring range
exceeded. Alcohol
residues that are still in the
mouth may cause
incorrect measuring
values.
Breath sample volume too
small.
(see order list, part number: 3700421) via
®
Troubleshooting
's instructions for use.
Remedy
Replace battery.
Blow into the mouthpiece
harder and continuously.
Allow the instrument to
cool down or become
warmer.
A wait time of at least
15 minutes after alcohol
was last orally consumed
must have elapsed (see
Page 8).
To repeat, press the
button.
Repeat breath sample.