Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Dimensions (W x D x
626.11 x 527.05 x 522.73 mm (24.65 x 20.75 x
H)
20.58 in.)
Enclosure
Rating: UL Type 4X (indoor), IEC 529 (IP 66) when air
purge is used and the door is closed and latched
Material: PC/ABS case, PC door, PC hinges and
latches, 316 SST hardware
Weight
22.4 kg (49.3 lb) dry, 28.9 kg (63.7 lb) wet with full
analyzer bottles
Mounting
Wall or bench
Protection class
I
Pollution degree
2
Installation category
II
Power requirements
100–240 VAC, 50/60 Hz, 100 VA
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Operating temperature
5 to 50 °C (41 to 122 °F); best performance: 5 to 35 °C
(41 to 95 °F)
Operating humidity
5 to 95% relative humidity, non-condensing
Maximum altitude
2000 m (6560 ft)
1
(approximately)
Table 1 General specifications (continued)
Specification
Details
Recorder outputs
Two 4–20 mA outputs applicable for recorders; load
impedance: 675 Ω maximum; 130 VAC isolation from
earth ground
Connection: 22 to 14 AWG wire, 22 to 20 AWG
recommended, twisted pair shielded wire
Alarm relay outputs
Two dry contact SPDT relays, not powered, each rated
at 5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Analyzer bottles
1 liter, high-density polyethylene with polypropylene
caps
Fuses
Two, 1.6 A, 250 V, 5 mm x 20 mm
External plumbing
Sample lines: 6.35 mm (0.25-in.) OD push-in fitting for
connections
poly tubing
Chemical drain (drain block): 12.7 mm (0.5-in.) ID barb
fitting for hose
Case drain: 25.4 mm (1-in.) ID for hose
Bottle enclosure drains: 9.53 mm (0.375-in.) ID hose
Air purge line: 6.35 mm (0.25-in.) OD push-in fitting for
poly tubing
Sample pressure
No minimum or maximum requirement
Sample flow rate
100–2000 mL/minute; 500–1000 mL/minute
recommended
Sample temperature
5 to 50 °C (41 to 122 °F)
Sample particle size
0.5 µm maximum; sample goes through a 0.5 µm filter
in the sample conditioning block before analysis
Number of sample
1; optional sample sequencing kit adds a second
streams
sample stream, sample sequence is programmable
Node-to-node distance
400 m (1320 ft) maximum between analyzers