Functional description
GB
On
Off
METHOD
Mode
METHOD
Zero
Test
METHOD
0.0.0
Zero
Test
METHOD
RESULT
Zero
Test
Zero
Test
42
Operation
Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key.
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the instru-
ment memorizes the last method used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw the
cap on and place the vial in the sample chamber making sure that the
marks are aligned.
Press the [ZERO/TEST] key.
The "Method" symbol flashes for approx. 8 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
marks are aligned.
Press the [ZERO/TEST] key.
(For Countdown / reaction period see page 43)
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
Repeating the zero:
Press the
[ZERO/TEST]
key for 2 seconds.
MD100_5a 08/2016