Methods
GB
Notes:
1. Complexed iron may be measured by increasing the development period until no further
colour development is seen. Very strongly complexed iron may not be included in the
measured iron. In this case the complexing agent must be destroyed by oxidation with
acid/persulphate followed by neutralisation to pH 6–9.
2. For total iron (suspended and dissolved), boil sample with acid/persulphate. Neutralise
back to pH 6– 9 making back up to original volume with distilled or deionised water.
3. When using KS61 (Ferrozine/Thioglycolate), high levels of molybdate will produce an
intense yellow colour.
In this case a reagent blank is required:
• Use two clean vials (24
• Mark one as blank for zeroing.
• Fill the blank with 10 ml sample.
• Add 10 drops KS63 (Thioglycolate).
• Close the vial tightly with the cap and swirl gently several times.
• Place the blank in the sample chamber making sure that the marks
• Press ZERO key.
• Remove the vial from the sample chamber.
• Fill a second clean 24 mm vial with 10 ml water sample (this is the sample).
Perform as described on page 46:
• Fill the vial with drops of the same size by holding the bottle vertically and squeeze
slowly:
10 drops KS61 (Ferrozine / Thioglycolate)
• Close the vial tightly with the cap and invert several times to mix the contents.
• Place the vial in the sample chamber making sure that the
• Wait for a reaction period of 5 minutes (Note 1).
(Countdown can be activated, see page 43)
• The method symbol flashes for approx. 3 seconds.
• The result is shown in the display in mg/l Iron.
Reagent / Accessories
KS61 (Ferrozine/ Thioglycolate)
KS63 (Thioglycolate Reagent)
Membrane-filter-set
MD100_5a 08/2016
mm Ø).
Form of reagent/Quantity
Liquid reagent / 65 ml
Liquid reagent / 65 ml
25 filter 0,45 µm
2 syringe 20 mL
are aligned.
marks are aligned.
Order-No.
56L006165
56L006365
366150
47