Electromagnetic Compatibility Information
Guidance and manufacture's declaration – electromagnetic emissions – for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The BM49 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer of the user of the BM49 blood pressure monitor
should assure that it is used in such and environment.
Emission test
Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
RF emission CISPR 11
Class B
Guidance and manufacture's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The BM49 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BM49 blood pressure monitor
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Power frequency (50Hz)
magnetic field
3 A/m
IEC 61000-4-8
The BM49 blood pressure monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The BM49 blood pressure monitor is suitable for use in all establishments, including domestic establishments and those directly con-
nected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Compliance
Electromagnetic environment – guidance
level
± 6 kV contact
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative
± 8 kV air
humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical com-
3 A/m
mercial or hospital environment.