Static Pressure Stability
Pressure change (cmH
O) at connection port at a pressure
2
setting of 10 cmH
O
2
Humidity
With ThermoSmart Breathing Tube
Maximum Humidity
Typical Humidity
With Standard Breathing Tube
Maximum Humidity
Typical Humidity
Electrical Ratings
Rated Supply Voltage
100 – 115 V
220 – 240 V
GAS TEMPERATURES:
NOISE LEVEL:
WATER CHAMBER VOLUME:
STANDARDS COMPLIANCE:
DATA RECORDING:
STORAGE AND TRANSPORT:
OPERATING CONDITIONS:
36 mg/L Body Temperature and Pressure Saturated (BTPS), 82% RH at 10 cmH
Humidity level 7 and Boost level high.
24 mg/L (BTPS), 90% RH at 10 cmH
32.24 mg/L, 73.21% RH at 10 cmH
17.97 mg/L, 85.88% RH at 10 cmH
Rated Current Input
1.27 A (1.43 A Max)
1.07 A (1.21 A Max)
Maximum = 100 °F (38 °C)
Sound Pressure Level <29 dBA
Average Sound Power Level <37 dBA
420 mL up to the maximum fill line
IEC60601-1: 1988 + A1 & A2; AS/NZS 3200.1.0: 1998
EN60601-1: 1990 + A1-A2; UL60601-1: 2003
The InfoUSB records up to 365 days of summary efficacy data, 7 days of
detailed efficacy data (AHI, Leak), Pressure and Leak data (50 Hz), as well
as cumulative data from first use. The device memory can also record up to
365 sessions of summary data or up to 200 sessions (400 hours) of detailed
efficacy data.
The device should always be stored and transported within the following
temperatures:
Transport:
14 to 140 °F (-10 °C to 60 °C)
Storage:
14 to 140 °F (-10 °C to 60 °C)
CAUTION: Ensure water is emptied from the Water Chamber before traveling.
Ambient Temperature: 41 to 95 °F (5 to 35 °C)
Humidity:
15 to 95% Relative Humidity
Altitude:
0 to 9,000 ft (0 to 3,000 m)
NOTE: Above 4,500 ft (1,500 m) the maximum operating pressure will be reduced.
A - 13
Auto/Premo
0.2
O, with Humidity level 4 and Boost level medium.
2
O with Humidity level 7 and Boost level high.
2
O with Humidity level 4 and Boost level medium.
2
Rated Supply Frequency
50 – 60 Hz/400 Hz
Novo
0.4
O, with
2
50 – 60 Hz