MEASURING PROCEDURES

8.8.1. Using the Instrument in a Single Phase System

The maximum voltage between B1 and B4 inputs is 600 V~ (CATII) / 350V~
phase – earth or 600V~ (CATIII) / 300 V~ phase to earth.
Do not measure voltages exceeding the limits prescribed by this manual.
Should you exceed the voltage limits you could damage the instrument
and/or its components or endanger your safety.
If possible, before connecting the instrument to the electrical equipment to
be tested take the power supply off the electrical equipment.
Phase
Neutral
B1
B2
Instrument connection in a single-phase system