Menu
Setting
13.Rounding
Automatically
Without rounding
20L
50L
100L
200L
500L
1.000L
14.–17.Exit
18.Language
Language:
+
name
Name:
19.Exit
20.LCD
Contrast: 24
display
21.Device
info
22.Test
current
Furthermore, devices connected on the relay contact will also be switched
on and/or off!
• Connected devices can be damaged (dry running).
• Operating media may leak.
23.Test
Relay 1 or
relay
Beeper =Off/On
SmartBox
Relay 2=Off/On
2 und 3
24.Reset
ESC
Restart
Factory settings
26.Exit
part no. 28 100 57 f
Disconnect the devices connected before test relay.
Only reconnect the devices again after test relay.
®
SmartBox
1 / SmartBox
Description
Default settings
minimal increments
Rounding increments in relation to the set volume
set with [+] / [-] keys
Press [Enter] to return to display mode
German, English, French
ESC
Name Tank 1: Suggested name
Letters can be changed with
Press [Enter] to return to display mode
Set the contrast of the LCD display
Software version: V6.00 (e. g.)
Serial number: Tank 1: SN=1234 (e. g.)
Offset + Gain: X0 = 4.05mA; B = 1268
Test function for the current mA value of the
probe: ADC = 7400 = 11.40mA
If level probe is not submerged, the value should
be close to 4mA. Tolerance range is 3.7..4.3mA.
Test function for the switch function of relay and
beeper
Exit this function without executing it.
Initialisation. The device software restarts and
keeps all device settings.
Complete reset of all parameters to the original
delivery status.
Press [Enter] to return to display mode
®
®
2 / SmartBox
[+] / [-] / [Enter]
[+] / [-] / [Enter]
[+] / [-] / [Enter]
15 / 20
3