Service programs
Test and adjustment program
®
Using the Testomat
808 Service Monitor software, you are able to
read data and to reset the meter readings (information on download-
ing the software can be found on page 6).
To do this, the software must be saved on a notebook that is con-
®
nected to the Testomat
interface. If the notebook doesn't have a serial interface, then please
use a USB 2 > serial adapter (Article no. 32286).
Access to the Test and Adjustment program
Press and hold key 1 (manual start) and switch the instrument on.
The POWER LED flashes.
Release key 1.
Alternatively
After switching on, hold down the reset key S1.
Then hold down key 1 (manual start) and release the reset key.
The POWER LED flashes.
Release key 1.
Operating the T808 Service Monitor Program
Open the "T808ServiceMonitor" program on your notebook by
double clicking.
Under Serial Port, select the COM port to which the Testomat
808 is connected.
Click Read to receive status information about the device in the
"Info-Data" section.
With Reset Runtime Head, you can set the counter of the pump
head operating hours meter to 0.
With Reset Rejection Counter, you can reset the counter for the
total number of rejections.
In the Output Window field, either the ADC value (for testing of
the yellowness index) or the value of the V level (for the clear wa-
ter comparison) will be displayed (see the description of the key
functions on pages 12-14).
To delete the Output Window field, click Clear.
808 with a null modem cable via the RS232
Service programs
®
15