Expected Charge Time
Battery charge time depends upon the level of discharge and ventilator
3
settings. Generally, one battery should be fully recharged within three hours
.
3
Two batteries should be fully recharged within five hours
. Charging two
SprintPack Batteries simultaneously will result in a quicker charge time.
The charging time for the SprintPack Batteries varies depending on the
power capability of the approved power source used.
NOTE
-
Charging time
The specified charging time for the SprintPack Batteries is
one full charge cycle from 0% to 95% capacity.
Batteries above 80% charge level - The Power Manager does not charge
a battery that is at or above 80% charge level if the battery had not been
removed from the Power Manager since the last use. This feature
maximizes battery cycle life.
Expected Number of Lifetime Cycles
A Cycle is a full discharge of a battery followed by a full recharge. The cycle
life (number of cycles) of each battery is determined by the depth of
discharge each battery continually encounters along with storage or
operating environment. A full cycle for each battery is from 95% to 0%
capacity (discharge) then recharged from 0% to 95% capacity (recharge).
Each SprintPack Battery has a lifetime of 400 cycles. After a battery has
undergone 400 cycles, replace it with Vyaire Medical SprintPack Battery, P/N
19444-001.
3
Utilize the LTV AC power adapter for best charge time results.
4-5
™
Operator's Manual – SprintPack
32328-001 Version A