C. Mechanical maintenance.
1. Adjusting the tension of the coarse focus control.
The tension adjustment control (Fig. 8) is located between the coarse focus knob (16) and the
arm (15). The coarse focus nov comes pre-adjusted by the manufacturer. The ideal tension point
is that which permits coarse focus knob to move as loosely as possible, without the stage (8)
moving down on its own.
a. To tighten the focussing controls of the coarse focus knob, the ring must be turned in an
anti-clockwise direction, as indicated by the arrow. To loosen it, the ring must be turned
clockwise.
Stage stopper
screw
Coarse focus
tension control
2. Adjusting the stage stopper.
The 40X and 100X (optional) objectives incorporate security measure in that the lens tip
retracts to avoid damage to the front of the lens should it come into contact with the slide.
An additional measure of security consists of an adjustable stopper on the ascending
movement of the stage. The stopper comes pre-adjusted by the manufacturer for standard
slides with cover slips 0.17mm thick. For observing other types of samples, however, it may
become necessary to readjust the stage stopper.
a. Loosen the stage stopper screw (Fig. 8) with a 2mm hexagonal key.
b. Focus on the sample using only the fine focus knob (5) with firstly the 4X objective, and
10X objective.
c. Tighten the stage stopper screw until firm enough for the stage is supported, and cannot
ascend further.
Fig. 8
-11-