RD to LNC
Adapter Cable
Reusable
INDICATIONS
The RD to LNC Adapter Cable has the same indications for use as the applicable sensors. Please see the compatible sensor's
directions for use for indications for use and prescribed information.
PCX-2108A
02/13
DESCRIPTION
The RD to LNC Adapter Cable allows RD SET™ sensors to be used with LNC series cables on Masimo® SET® devices or to
connect to devices with conventional Nellcor Technology.
The RD to LNC Adapter Cable has been verified using Masimo SET technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET oximetry or licensed to use
Masimo sensors.
WARNINGS
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the
cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
INSTRUCTIONS
A) Attaching the RD to LNC Adapter Cable to a Patient Cable
1. Refer to Fig. 1. Properly orient and insert the connector completely into the patient cable connector.
2. Refer to Fig. 2. Completely close the protective latch.
B) Disconnecting the RD to LNC Adapter Cable from a Patient Cable
1. Refer to Fig. 3. Lift the protective cover to gain access to the connector.
2. Refer to Fig. 4. Pull firmly on the connector to remove from the patient cable. To avoid damage, pull on the connector,
not the cable.
CLEANING
1. Remove the sensor from the patient and disconnect it from the adapter cable .
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad .
3. Wipe all surfaces of the cable
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable .
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
CAUTION:
• Do not immerse the connector on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
Non-Sterile
3
en
9182D-eIFU-0916