GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS

The equipment AP 500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment AP 500 should assure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations / Flicker emissions
IEC 61000-3-3

GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY

The equipment AP 500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment AP 500 should assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-5
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/60 HZ)
magnetic fi eld
IEC 61000-4-8
TECNIWORK S.p.A. - PODOSCOPE AP500
PodoManISTRUZ_A4_x.indd 20
COMPLIANCE
Group1
Class B
Compliant
Compliant
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
±2 KV contact
±2 KV contact
±4 KV contact
±4 KV contact
±6 KV contact
±6 KV contact
±2 KV air
±2 KV air
±4 KV air
±4 KV air
±8 KV air
±8 KV air
±2 KV for power
±2 KV for power
supply lines
supply lines
±1 KV differential mode
±1 KV differential mode
±2 KV common mode
±2 KV common mode
<5 % U
<5 % U
T
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0,5 cycle
for 0,5 cycle
40 % U
40 % U
T
T
(60 % dip in U
) for cycle
(60 % dip in U
T
70 % U
70 % U
T
T
(30 % dip in U
) for cycle
(30 % dip in U
T
<5 % U
<5 % U
T
T
(>95 % dip in U
) for 5 s
(>95 % dip in U
T
3 A/m
3 A/m
- 20 -
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
The equipment "AP 500" must emit elec-
tromagnetic energy in order to perform its
intende function. Nearby electronic equip-
ment may be affected.
The "AP 500" is suitable for use in all esta-
blishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buil-
dings used for domestic purposes.
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Floor should be wood, concrete or ceramic
tile. If fl oors are covered with synthetic
material, the relative humidity should be at
less 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
)
commercial or hospital environment.
T
If the user of AP 500 requires continued
operation during power mains interruptions,
it is reccomended that the AP 500 be powe-
) for cycle
red from an uninterruptible power supply or
T
a battery.
) for cycle
T
) for 5 s
T
Power frequency magnetic fi eld should
beat levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
doc.cod.IS-05-2 Ed.23/11/16