Configuración de los parámetros operativos
Tabla 5-5
Alarmas de estatus y niveles de severidad
Mensaje del comunicador
Código
de alarma
Mensaje de ProLink II
A001
EEprom Checksum Error (Core Processor)
(E)EPROM Checksum Error (CP)
A002
RAM Test Error (Core Processor)
RAM Error (CP)
A003
Sensor Not Responding (No Tube Interrupt)
Sensor Failure
A004
Temperature sensor out of range
Temperature Sensor Failure
A005
Input Over-Range
Input Overrange
A006
Transmitter Not Characterized
Not Configured
A008
Density Outside Limits
Density Overrange
A009
Transmitter Initializing/Warming Up
Transmitter Initializing/Warming Up
A010
Calibration Failure
Calibration Failure
A011
Excess Calibration Correction, Zero too Low
Zero Too Low
A012
Excess Calibration Correction, Zero too High
Zero Too High
A013
Process too Noisy to Perform Auto Zero
Zero Too Noisy
A014
Transmitter Failed
Transmitter Failed
A016
Line RTD Temperature Out-Of-Range
Line RTD Temperature Out-of-Range
A017
Meter RTD Temperature Out-Of-Range
Meter RTD Temperature Out-of-Range
A020
Calibration Factors Unentered
Calibration Factors Unentered (FlowCal)
A021
Unrecognized/Unentered Sensor Type
Incorrect Sensor Type (K1)
A029
Internal Communication Failure
PIC/Daughterboard Communication Failure
A030
Hardware/Software Incompatible
Incorrect Board Type
Manual de configuración y uso
Severidad
predeterminada
Configurable
Fault (fallo)
No
Fault (fallo)
No
Fault (fallo)
Sí
Fault (fallo)
No
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
No
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
No
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
Sí
Fault (fallo)
No
Fault (fallo)
No
Fault (fallo)
No
Afectado por el
timeout del último
valor medido
No
No
Sí
Sí
Sí
No
Sí
No
No
No
No
No
No
Sí
Sí
No
No
No
No
57