Instruction Manual DIGIMON-SE / DIGIMON4
Calibrating pressure sensors
A correct calibration requires a certified monitor.
(e.g. REF-CLASS-GAUGE P/N 4682293)
1. Switch off the DIGIMON
2. Press down the
button.
The display shows "Password" 000
3. Press the
4. Press the
The display shows "P-call"
5. Press the
400 psi should be displayed on the LP side display
For a correct calibration, there should be no pressure or vacuum in the
DIGIMON.
6. Press the
pressure (400psi).
7. Press the
8. Wait 3-5 seconds until the pressure is stable and press the
again.
The display shows "0000"
9. Pressurise the DIGIMON with a calibration pressure of 400psi (27.58 bar).
When the pressure is stable, press the
The display shows "LP End"
10. The calibration of the low pressure side has been completed.
11. Press the
from Point 6 or press the
LIGHT
button and switch on the device using the
+
-
R
R
/
button to select Code 009.
SET
button to exit.
-
R
button and then the
+
R
button to select the low pressure side's calibration
SET
button to exit.
-
R
button to calibrate the high pressure side (HP) Repeat
ON/OFF
SET
button to calibrate.
SET
button to confirm.
button to end the calibration.
English
ON/OFF
SET
button
17