Power
Nominal 24 V DC at max. 550 Ohm
• Maximum 30 V DC
• 4 to 20 mA
Approvals
• General
CSA
• Radio
Europe (R&TTE), FCC, Industry Canada
• Hazardous Intrinsically Safe
Non-incendive
• Marine
Lloyd's Register of Shipping
ABS Type Approval
The use of approved watertight conduit hubs/glands is required for Type 4X /
Note:
NEMA 4X, Type 6 / NEMA 6, IP67, IP68 (outdoor applications).
1.
Wiring Drawing: Intrinsically Safe (FM )
See
Wiring Drawing: Intrinsically Safe (CSA)
or
2.
Wiring Drawing: Non-incendive (FM)
See
7ML19985QP81
, FM, CE
US/C
(Europe)
(US/Canada)
(Australia)
(International) IECEx TSA 04.0020X Ex ia IIC T4
(US)
on page 2 of Appendix A, for drawing number 23651611,
on page 3 of Appendix A, for drawing number 23651621.
on page 1 of Appendix A, for drawing number 23650537.
SITRANS Probe LR – QUICK START MANUAL
ATEX II 1 G EEx ia IIC T4
1
FM/CSA
: (barrier required)
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Class III
ANZEx Ex ia IIC T4
(Tamb = –40 to + 80 °C) IP67, IP68
2
FM
:
Class I, Div. 2, Groups A, B, C, D
T4
T5
Page EN-3