NOTE: If there are three beeps, this indicates a low battery condition for the 9-volt
battery inside the cabinet of the Eclipse. please contact your home care provider and
request that the 9-volt battery be replaced.
NOTE: Please allow longer warm up times for cold weather start-ups.
the eclipse has an oxygen concentration status indicator (OCSi) built in to the device. the OCSi
continually monitors the oxygen output of the device and incorporates a green light (normal),
yellow light (alert), red light (alarm) and a buzzer. these indicators will all illuminate during the
power on self-test at start-up. After initially powering ON the device, please allow up to five (5)
minutes for the device to reach its performance specifications. When only the green indicator
light is illuminated, the device has reached its performance specifications.
Step 3: Connect the Oxygen Supply Tubing, or Nasal Cannula to the
Oxygen Outlet.
replace the cannula and oxygen supply tubing regularly, as recommended by your home care
provider. Consult your home care provider for cannula and supply tubing cleaning and replace-
ment information.
Step 4: Select the Flow mode
Press the Flow mode button to select the desired mode. Pressing this button repeatedly will
toggle you back and forth between Pulse Dose mode and Continuous Flow mode. the proper
mode and proper setting should be prescribed for you by your physician.
CONTINUOUS FlOW mODE OpERATION:
When operating in the Continuous Flow mode, a continuous supply of oxygen measured in
liters per minute (lPm) will flow through your tubing and nasal cannula.
pUlSE DOSE mODE OpERATION:
When operating in Pulse mode, a bolus of oxygen, measured in milliliters (ml), is delivered
upon inspiration. DO NOt use a humidifier in the Pulse Dose mode.
Secure a prescription for pulse dose from your physician with consideration to your needs at
rest, during exercise and when traveling at altitude.
When operating in pulse dose mode, the eclipse 3 supplies a continuous flow of oxygen when
a breath is not detected. the device will continually check for an inspiratory effort every 15
seconds. Once an inspiration is detected, the device reverts to measured bolus delivery. to exit
19
Pulse Dose mode and engage Continuous Flow mode, push the flow mode button.
Personal Ambulatory Oxygen System Users Manual