pUlSE DOSE mODE AND autoSAT FEATURES
Your eclipse has a feature, called autoSAt
to 40 breaths per minute (see max breath rate chart below). As your breath rate increases, ie.
during ambulation, the autoSAt feature servo-controls the device delivering the set bolus size.
autoSAt technology is proprietary to SeQual and is utilized during Pulse mode only.
Pulse
Dose
Setting
1.0
2.0
3.0
4.0
5.0
6.0
-
-
-
NOTE: bolus volume decreases as breath rate exceeds published range.
WARNING:
DO NOT USE SUPPLY TUBING OR CANNULA EXTENSIONS THAT ARE MORE THAN 50' (15.2M) IN LENGTH
IN THE CONTINUOUS FLOW MODE.
PULSE DOSE MODE SETTINGS MUST BE DETERMINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR
NEEDS AT REST, DURING EXERCISE AND WHEN TRAVELING.
DO NOT EXCEED 7 FEET (2.1M) IN LENGTH OF SUPPLY TUBING OR CANNULA TUBING WHEN OPERATING
IN PULSE DOSE MODE IN ORDER TO ADEQUATELY DETECT ADEQUATE INSPIRATORY EFFORT.
AS WITH ALL OXYGEN-CONSERVING TYPE DEVICES, THE ECLIPSE MAY NOT BE ABLE TO DETECT ALL
INSPIRATORY EFFORTS IN PULSE MODE. IF THE ECLIPSE DOES NOT RESPOND TO YOUR INSPIRATION
EFFORTS, ASk YOUR HOME CARE PROVIDER TO CHECk THE SENSITIVITY SETTING.
Step 5: ADjUST ThE FlOW SETTING TO ThE pRESCRIbED lEvEl
using the Flow Setting buttons, adjust the flow setting to the prescribed setting supplied by
your physician.
WARNING:
DO NOT CHANGE FROM PRESCRIBED FLOW SETTING WITHOUT CONSULTING WITH A QUALIFIED
CLINICIAN OR PHYSICIAN.
20
Personal Ambulatory Oxygen System Users Manual
®
AC Power Supply
Bolus Size
Cartridge (Battery)
(± 15%) mL
16
32
48
64
80
96
128
160
192
that delivers a consistent-sized bolus of oxygen up
and Power
Max Breath
Rate
40
40
40
40
37
31
23
18
15
DC Power
Supply
Max Breath
Rate
40
40
40
31
25
20
15
12
10