3. You activate the treatment parameters
and functions with the four parameter
keys below the display. The symbols
above the four parameter keys indicate
the assigned parameters and functions.
This is what happens when you press
one of the parameter keys to select a
parameter:
• The corresponding symbol appears
on the display in a larger format.
• The set value is displayed.
• The symbol above the parameter key
appears in reverse video.
4. With the + / - keys (plus/minus) you
change the displayed value. When you
press and hold the key, the value will
change at a higher rate.
Some of the (special) functions can only
be enabled and disabled. This is done by
pressing the corresponding parameter
key or with the + / - keys. Active param-
eters are identified with a check mark in
the circle next to the symbol.
5. Having programmed all parameters,
press the STOP key to save the values.
6. Then press the START key: the
ARTROMOT®-E2/-E2 compact checks the
set values, moves to the position halfway
between the set extension and flexion
values as well as to the set supination
and pronation values and stops.
7. Press the START key again to start
therapy.
The CPM device will now begin with the
set physiological motion pattern in the
synchronized mode. The motor with
the larger range of motion operates at
the programmed speed, while the speed
of the second motor is adapted. This
ensures that both motors will reach their
target positions at almost the same time.
After activation of the START key in the
non-synchronized mode the motors
will perform random movements, each
motor reversing after reaching the maxi-
mum values. The programmed speed is
valid for both motors.
Note!
− Refer to section 5.3 for a description
of the parameters.
− To view the set parameter values,
press the corresponding parameter
key. Before, however, you have to
press the STOP key and access the
correct menu level.
− To prevent accidental changes of the
parameter settings, you can lock the
keys. To do so, simultaneously press
keys + and – for approx. 3 seconds.
Press both keys again for approx. 3
seconds to unlock.
− Selecting the "new patient" function will
automatically delete the data on the pa-
tient chip card. When you have finished
programming the unit and press the
STOP key, the settings will automati-
cally also be saved to the patient chip
card.
− Emergency stop function: When any of
the keys is pressed during therapy, the
ARTROMOT
-E2/-E2 compact switches
®
immediately off.
– When the START key is pressed again
in the synchronized mode and in the
non-synchronized mode, the carriage
will reverse its direction.
Patients with a
programmed chip card
• First complete the mechanical
adjustments.
• Then insert the chip card (the patient is
not yet positioned on the CPM device).
• Press the START key: the device will
move to the starting position as speci-
fied by the parameters stored on the
chip card and stops.
• Position the patient on the CPM device
and press the START key to initiate
therapy.