9. IEC 60601-1-2:2001
The ARTROMOT®-E2/-E2 compact is
subject to particular precautions regard-
ing electromagnetic compatibility (EMC).
The device must be installed and put into
service strictly in compliance with the EMC
directives put forth in the accompanying
documents.
Portable and mobile RF communication
systems may affect the
ARTROMOT®-E2/-E2 compact.
The ARTROMOT®-E2/-E2 compact should
not be used adjacent to or stacked with
other equipment. If adjacent or stacked use
is necessary, the
ARTROMOT®-E2/-E2 compact should be
observed to verify normal operation in the
configuration in which it will be used.
If you detect damage or malfunctions that
may impair the safety of the patient or
9.1 Electromagnetic emissions
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The ARTROMOT®-E2/-E2 compact is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to ensure that the ARTROMOT®-E2/-E2 compact is
used in such an environment.
Emissions test
RF emissions
to CISPR 11
RF emissions
to CISPR 11
Harmonic emissions to IEC
61000-3-2
Voltage fluctuations/flicker
emissions to IEC 61000-3-3
of the operator, have the device repaired
before using it.
If it is necessary to replace assemblies or
cables, only the manufacturer's original
parts must be used to ensure continued
compliance with EMC requirements after
repair. This requirement applies to the pow-
er supply unit, cables and cable lengths,
drive unit consisting of the motor and the
control system, the programming unit incl.
the coiled cable and the connector.
Compliance
Electromagnetic environment - guidance
Group 1
The ARTROMOT®-E2/-E2 compact uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The ARTROMOT®-E2/-E2 compact is suitable for
use in all establishments, including domestic
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
not applicable
not applicable
84