Press the
confirm that the unit is switched off.
9

Troubleshooting

How to proceed in case of malfunctions and errors

Error messages disappear immediately from the display as soon as the knee plate or foot
lever is released or the fault is eliminated. This means there are no waiting times for re-
starting.
Exception: If the table control is operated without a foot switch, all error messages must
be reset by simultaneously pressing and holding the touch fields O + + . The control
unit can be completely reset by disconnecting from the power supply.
If a fault cannot be solved with the above procedure, please contact an authorised service
partner or Schick directly.

Fault and error messages

The
potential faults and display them as fault codes on the control satellite. If a malfunction
occurs, this is indicated by a red-lit display of the control satellite, a red flashing control
unit and an error message in the format "Exxyy".
D21518
Original

Switching off

symbol on the operating unit for more than two seconds. Three short beeps
control unit features intelligent control electronics that can detect
Example of error message: E0804 error