Product information
Maintenance
After 2,000 hours of operation, KINETEC C.E.M. requires a few lubrication and maintenance operations (lubrication of the joints, pointer stops and ball screws).
The need for maintenance is indicated by display of the message SERV. MOTOR when the system is switched on.
Despite that warning, you can continue to use your KINETEC C.E.M. by pressing START, but you should contact your nearest KINETEC technician to have the
maintenance operations conducted as soon as possible.
• A motor running time counter is available. Simultaneously press keys
RESET TIME 924H
Reset: limit low
Cleaning
Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal
components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
Recommendations to obtain a maximum hygiene of the pads.
• Sterilization of the pads (if necessary):
Sterilizer at 134°C during 18 minutes.
• Disinfection of the pads:
Washing at 30°C with use of a disinfection solution during the rising cycle.
Technical specifications
Product:
Weight: 4,6kg (10,1lb)
Unit dimensions: 20x18x62cm
(7,9x7,1x24,41inches)
Angular limits: -10 to 135°
Speed: from 50° to 140°/min
Patient sizing:
from 4'7" to 6'7"
1,40 to 2m
Troubleshooting
A spare parts list and technical catalog are available to you on request from your KINETEC distributor.
If, after connecting the power supply cable to the power supply and switching on KINETEC C.E.M.:
• The display does not indicate any information:
- Check that the electrical socket is live using another device.
- Replace the fuse(s) (33) of the connector with fuses of the same type and caliber:
2 fuses T 750 mA 250V (6.3 x 32) (KINETEC order: 4610007434).
- If the display still does not indicate any information, contact your nearest KINETEC technician.
• Your KINETEC C.E.M. does not work and the display indicates 50 STOP 25 115,
Press START again.
Your KINETEC C.E.M. still does not function: Contact your nearest KINETEC technician.
• Your KINETEC C.E.M. does not function and the display indicates:
SERVICE D1: angle measurement function failure,
or SERVICE D2: no movement,
or SERVICE D3: abnormal consumption,
or SERVICE D4: motor connection failure,
Contact your nearest KINETEC technician if the same message is displayed after having switched the
device off, then on, and started it by pressing START.
GB
16
this counter can be reset by pressing the key
Electrical:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm
(KINETEC order: 4610007434)
and
, the displays shows
.
Environment
- Storage/transport conditions:
Temperature: -40 to 160°F / -40 to 70°C
Relative humidity: up to 90%
- Operating conditions:
Room temperature: 50 to 105°F / 10 to 40°C
Relative humidity: up to 80%.
33