Performing a Control Solution Test
You will need:
Meter
1
WARNING:
When doing the control solution test, you have to mark it so that the test result
will NOT be stored in the memory. Failure to do so will mix up the blood glucose
test results with the control solution test results in memory.
Test Strip
2
Control Solution
3
1. Insert the test strip to turn on the meter
Wait for the meter to display "
" ".
2. Press M to mark this test as a control
solution test
With "
" displayed, the meter will store
the test result in memory under "
you press the M button again, the "
will disappear and this test is no longer a
control solution test.
3. Apply control solution
Shake the control solution vial thoroughly
before use.
Squeeze out a drop and wipe it off, then
squeeze out another drop and place it
on the tip of the vial cap. Hold the meter
on an angled position and ensure the
" and
". If
"