Temperature Measurement (Optional) - Gima OXY 110 Guía De Uso

Ocultar thumbs Ver también para OXY 110:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

 The SpO
sensor cannot be immersed into water, liquid or cleanser.
2
 The SpO
sensor can be repeatedly used. Please clean and disinfect before reuse.
2
 Connector with the label "SpO
connector with the label "TEMP" can only be connected with the temperature
probe.

3.2 Temperature Measurement (optional)

The infrared temperature probe is a delicate transducer. To operate please follow
these steps and procedures. Failure to accurately operate may cause damage to
the probe.
The infrared temperature probe is as shown in figure 3.2.
Please place the infrared temperature probe in a stable ambient temperature for
30 minutes before taking a measurement.
Measuring tip of
temperature probe
Temperature
Probe Cover
Small screen
Power on/off key
Battery cover
Figure 3.2 the infrared temperature probe
Operation procedure:
1. Connect the infrared temperature probe to the connector on the upper side of
device marked with "TEMP".
2. When the screen shows as the big display screen in figure 3.2 and the tempe-
rature unit °C is blinking, the user can begin to take the measurement.
3. Insert the tip of the temperature probe into the earhole and press the measuring
key to start the measurement. A short beep means the measurement has fini-
shed and the result will be displayed on the big display screen on temperature
probe and the display screen of the Oximeter.
Note:
• If the temperature probe detects a hardware failure, the display screen on the
infrared temperature probe will show "Err" and will not enter into measurement
mode.
ENGLISH
" can only be connected with SpO
2
probe, and
2
Measuring key
Big display screen
46

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido