Controls, Alarms And Indicators; Pump Controls; Static Mode; Alarm Mute - Arjohuntleigh Nimbus 4 Instructiones De Uso

Ocultar thumbs Ver también para Nimbus 4:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 119

4. Controls, Alarms and Indicators

Run/Standby
Button
Run Mode
Indicator
Static Mode Control
Pump Controls The pump front panel has the following controls:
Run/Standby Button

Static Mode

Alarm Mute

Mute Control
& Indicator
Indicator
Indicator
& Indicator
Press the Run/Standby button to put the pump into the
Run mode; the Run indicator will change to green.
To put the pump into Standby, press and hold the Run/
Standby button for approximately 3 seconds; this
prevents accidental operation. The Run indicator will be
extinguished.
Selects the operating mode, either Static or Dynamic.
When the pump is first powered up, its default setting is
Dynamic mode. To switch to Static mode, press and
hold the Static Mode button for a minimum of 3
seconds. Static mode is confirmed by the illumination of
the button's yellow indicator.
To return to Dynamic mode, press and hold the button
for a minimum of 3 seconds. Dynamic mode is
confirmed by the extinguishing of the button's yellow
indicator.
An Alarm Mute button is provided to silence warning
sounds during an alarm condition. Press and hold the
Alarm Mute button for a minimum of 3 seconds to mute
any alarms.
Alarm Mute is confirmed by the illumination of the
button's yellow indicator.
The Alarm Mute button does not operate in a Power
Fail condition.
Alarm
High Pressure
Indicator
Wait
Service
Indicator
Power Fail
Indicator
10
Low Pressure
Indicator
Comfort
Control
Pump Fault
Indicator

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Nimbus professional

Tabla de contenido