Calibration Check - Vyaire Micro I Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 172

Calibration Check

The Micro I is calibrated to read in litres at body temperature,
barometric pressure saturated with water vapour (BTPS).
The calibration should remain stable indefinitely, unless the
transducer is physically damaged, and the unit should not require
re-calibration. However, to ensure the correct functioning of the unit
we do recommend that a calibration check is performed after each
patient or when the transducer was removed for cleaning.
To perform a calibration check select calibration check option from
the setting menu and the display will show:
Connect a 3L syringe to the Micro I with the minimum of adapters
and inject the syringe volume into the transducer evenly, without
pausing. When complete the display will show:
The acceptable calibration accuracy is +/- 3.5%. If the calibration
accuracy is outside of this range then the tick will be replaced by a
cross. If this happens then check the syringe and the connections for
leaks and repeat the check. If the Micro I is repeatedly outside the
34

Publicidad

Tabla de contenido
loading

Productos relacionados para Vyaire Micro I

Tabla de contenido