Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Micro I is intended for use in the electromagnetic environment specified
below. The customer or the user of the Micro I should assure that it is used in
such an environment.
Immunity Test
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
IMMUNITY to
proximity fields
from RF wireless
communications
equipment IEC
61000-4-3
IEC 60601 test level
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
28 V/m
450 MHz, 50% PM at 18 Hz
810 MHz, 50% PM at 18 Hz
870 MHz, 50% PM at 18 Hz
930 MHz, 50% PM at 18 Hz
1720 MHz, 50% PM at 217 Hz
1845 MHz, 50% PM at 217 Hz
1970 MHz, 50% PM at 217 Hz
2450 MHz, 50% PM at 217 Hz
27 V/m
385 MHz, 50% PM at 18 Hz
9 V/m
710 MHz, 50% PM at 217 Hz
745 MHz, 50% PM at 217 Hz
780 MHz, 50% PM at 217 Hz
5240 MHz, 50% PM at 217 Hz
5500 MHz, 50% PM at 217 Hz
5785 MHz, 50% PM at 217 Hz
44
Compliance Level
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
28 V/m
450 MHz, 50% PM at 18 Hz
810 MHz, 50% PM at 18 Hz
870 MHz, 50% PM at 18 Hz
930 MHz, 50% PM at 18 Hz
1720 MHz, 50% PM at 217
Hz
1845 MHz, 50% PM at 217
Hz
1970 MHz, 50% PM at 217
Hz
2450 MHz, 50% PM at 217
Hz
27 V/m
385 MHz, 50% PM at 18 Hz
9 V/m
710 MHz, 50% PM at 217
Hz
745 MHz, 50% PM at 217
Hz
780 MHz, 50% PM at 217
Hz
5240 MHz, 50% PM at 217
Hz
5500 MHz, 50% PM at 217
Hz
5785 MHz, 50% PM at 217
Hz