The Micro I has an essential performance – When verified with a 3 litre
syringe the readings remain within a tolerance of +/-3.5%, and unit
firmware must not cease operating. In the unlikely event of a Fast Transient
/ ESD event occurring, the device should be reset and located away from the
source of interference.
Note: In the very unlikely event of a high ESD event occurring to
the device connector, the device could be damaged.
WARNING:
The Micro I should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use with other equipment
is necessary, the Micro I and the other equipment should be observed /
monitored, to verify normal operation in the configuration in which it will be
used.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Micro I is intended for use in the electromagnetic environment specified
below. The customer or the user of the Micro I should assure that it is used in
such an environment
Emission Test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic
Emissions
IEC 61000-3-2
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class [B]
[Not Applicable]
[Not Applicable]
Electromagnetic Environment
-Guidance
The Micro I uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The Micro I is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purpose.