Tab. 1
FLANGE
ISO 63
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TPS-mini cannot be fixed by means of its base.
NOTE
For installation of optional accessories, see "Technical Information".
WARNING!
The System must be installed in a way that allows an easy interruption of the
line voltage.
WARNING!
Power supply cord: The correct cable for electrical wiring is a three wires
(Ph+N+Earth) cable. The wire section has to be at least 0.75 mm
TPS-mini User Manual / 87-901-003-01 (E)
FIXING DEVICE
M10 clamps
M8 claws
Instructions for Use
Installation
N.
FIXING TORQUE
4
22 Nm
4
11 Nm
16
(AWG18).
2
259/318