17
Technical Information
Technical Specification
Technical Specification
Tab. 3
Pumping speed
Compression ratio
Base pressure * (with
recommended forepump)
Inlet flange
Foreline flange
Rotational speed
Start-up time
Recommended forepump
Operating position
Cooling requirements
Operating ambient temperature
Coolant water
Bakeout temperature
Noise level
Pwr supply (mains):
Protection fuse
224/282
Technical Specification
Input voltage:
Input freq.:
Max input pwr:
Stand-by pwr:
Max oper. power:
TV 301 Navigator User Manual / 87-900-946-01
With CFF 8" or ISO 160:
N
: 280 l/s
2
He: 230 l/s
H
: 210 l/s
2
N
: >2 x 108
2
He: 1 x 105
H
: 2 x 104
2
mechanical: < 2 x 10-10 mbar (< 1.5 x 10-10 Torr)
CFF 8" O.D.
ISO 160
CFF 6" O.D.
ISO 100
KF16 NW (KF25 – optional)
56000 rpm
< 3 minutes
mechanical: Agilent DS 102
mechanical: Agilent DS 302
dry pump: Agilent SH 110
any
Natural air convection, Forced air or water optional
+ 5 ° C to + 35 ° C
flow: 200 l/h (0.89 GPM)
temperature: + 10° C to + 30° C
pressure: 3 to 5 bar (45 to 75 Psi)
120° C at inlet flange max. (CF flange)
80° C at inlet flange max. (ISO flange)
≤ 45 dB (A) at 1 meter
100 - 240 Vac
50 - 60 Hz
300 VA
10 W
150 W with water or air cooling
4 A
With CFF 6" or ISO 100:
N
: 250 l/s
2
He: 220 l/s
H
: 200 l/s
2
(M)