Operating Instructions SIROEndo
4
A
60 78 104 D 3488
D 3488.201.01.07.02 11.2009

Installation

4.1

Preparations

1.
Cut through the adhesive tape and open the cardboard box.
2.
Remove the SIROEndo and its accessories from the cardboard box in
their original packaging and lay everything down on a flat surface.
3.
Before installing the SIROEndo, inspect everything for damage. Any dam-
age which may have occurred during transport must be reported to the
dealer within 24 hours after receipt of the SIROEndo.
4.
Check the contents of the cardboard box for completeness as described
in section: Scope of supply on page 9.
CAUTION !
This product is intended only for use by trained dental personnel.
Do not install SIROEndo near direct or indirect sources of heat.
Do not insert inappropriate objects in the drive shaft of the micromotor.
4.2

Installation on the support tube

1.
Determine the required thickness of the hard rubber collars (A).
To do this, place the end of the supplied ruler on the support tube in such
a way that the black dot is located at the position where you would like to
install the SIROEndo. Holding the first end firmly against the support tube,
wrap the another end around the support tube.
Example: The max. internal adapter diameter minus the measured diam-
eter of the support tube equals the thickness which must be compensated
for by the hard rubber collars.
2.
Lay the hard rubber collars included in delivery (A) in the
adapter (11 or 12).
Fasten the adapter (11 or 12) for the support arm to a support tube of your
11/12
treatment center with four screws (3 mm Allen key).
The screws must be tightened firmly!
CAUTION
Make sure that the support tube is suitable for fastening the SIROEndo.
Check the adapter to make sure that it is firmly seated on the support tube.
i
NOTE
The installation of the wall adapter is described in a separate set of installation
instructions.
e.g.: Adapter dia. = 56 mm
Lamp support dia. = 54 mm
Thickness of hard rubber collar = 2 x 1 mm
4 Installation
11