The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TV 1001 Navigator ca be installed in any position.
The TV 1001 Navigator cannot be fixed by means of its base.
NOTE
The TV 1001 Navigator belongs to the second installation (or overvoltage)
CAUTION!
category as per directive EN 61010-1. Connect the device to a mains line that
satisfy the above category.
The TV 1001 Navigator has Input/Output and serial communication connectors
that must be connected to external circuits in such a way that no electrical part
is accessible.
Be sure that the insulation of the device connected to the TV 1001 Navigator is
adequate even in the case of single fault as per directive EN 61010-1.
TV 1001 Navigator User Manual /
87-900-945-01 (I)
Instructions for Use
Pump fixing
233/312