CABLES USED FOR THE TEST
Never disconnect the test leads while the word "Measuring" is being
displayed.
4.1.1.1.
Procedure to cancel calibration parameters and cal symbol
To cancel the calibration parameters (and
the symbol CAL) it is necessary to perform a
calibration procedure with a resistance
greater than 5Ω (for example with the test
leads disconnected). When a cancellation is
completed this screen is displayed first,
followed by the screen below:
The screen alongside is displayed for 2
seconds, after which the instrument emits a
long beep and then displays the default
screen relative to the LOWΩ test under
AUTO mode without the symbol CAL.
If the cables are changed or extended be sure and
recalibrate the instrument.
test, if the resistance value free of calibration (that
is the resistance value less the calibration offset
value) is negative, the symbols
blinking CAL are displayed (refer to 5
paragraph 4.1.3).
WARNING
Ω
LOW
ν> 5
Ω
cal
Measuring
Ω
LOW
es
cal
During a continuity
ν
as well as
th
screen
Message >5Ω:
means that the
instrument
detected a
resistance greater
than 5Ω therefore
it will proceed
with Reset
procedure.
Message rES:
means that the
instrument
cancelled
(RESET) the
calibration
parameter