7.
Press the START/STOP key. The instrument starts the test.
At the end of the test the instrument
displays the values alongside.
Earth resistivity value
Never disconnect the test leads while the word "Measuring" is being
displayed.
SAVE
SAVING:
Note: If START/STOP key is pressed & held, the instrument takes more consecutive
tests.
8. When a new value is acquired the instrument emits a short sound and calculates the
new average value.
9.
DISP
Press DISP key to show how many measurements are in average value
calculation and the measuring mode.
10.
Press CLR key if you want to cancel the medium value of the resistance
CLR
and the no. of measurements which are included in the calculation.
Note: Any test with result over 1999Ω is not inserted in the calculation of medium value.
st
Example: 1
measurement (D=1):
main display: 6,6Ωm
secondary display: 6,6Ωm
nd
2
measurement (D=1):
WARNING
The tests can be stored by pressing the SAVE key twice
(refer to paragraph 5.1).
ρ
36.5
Ω
m
2
0.93 Ω