17
Technical Information
RS 232/RS 485 Communication Description
N.
108
109
110
111
120
121
122
123
124
125
126
130
200
201
268/300
Read/Write
Datatype
R/W
N
W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
L
Reserved to Agilent service
R/W
L
R/W
N
Reserved to Agilent service
R
N
R
N
TV 301SF Navigator User Manual / 87-900-957-01 (H)
Description
Baud rate
Pump life/ cycle time/ cycle
number reset
Interlock type
(default = 1)
Analog output type: output
voltage signal proportional to
frequency or power
Rotational frequency setting
(Hz)
Maximum rotational
frequency in Hz (active only
in Stop condition)
Set vent valve on/off
(on = closed)
Set the vent valve
operation
Vent valve opening
delay
(expressed in 0.2 sec)
Pump current in mA dc
Pump voltage in Vdc
Admitted Values
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write
'1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
150 to 963
(default = 963)
150 to 963
(default = 963)
On = 1
Off = 0
(default = 1)
Automatic = 0
(see note 1.)
On command =
1 (see note 2.)
0 to 65535
(corresponding
to 0 to 13107
sec)