Inlet Screen Installation
Figure 25
The inlet screen mod. 969-9302 prevents the blades of the pump from
being damaged by debris greater than 0.7 mm diameter. The inlet
screen, however, does reduce the pumping speed by about 10 %. The
inlet screen is fitted in the upper part of the pump, as shown in the
figure.
Figure 26
TV 301SF Navigator User Manual /
The screen can be mounted on each pump.
87-900-957-01 (H)
Technical Information
Inlet Screen Installation
17
271/300