Network Connector (Rs 485) - Agilent Technologies Turbo-V 3K-G Manual De Instrucciones

Publicidad

Interconnections
The following figure shows the MoniTorr's interconnections.

Network Connector (RS 485)

Figure 30
In order to have the IP-54 environment protection use the
appropriate connector provided.
To wire the MoniTorr network please connect all the A+ terminal
toghether and the same for B- and Gnd terminals. In order to
guarantee a well noise immune communication if you create the
network by yourself please remember to insert at the first and at the
last connectors of the RS485 cable two terminator resistors (120 Ohm
½ watt) and to use only shielded cable. In any case if you prefer
Agilent provides the right cable adapted for your system: please
contact our customer service.
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
Network connector (RS-485)
Technical Information
MoniTorr