User performance results for glucose concentrations between
48 mg/dL and 553 mg/dL.
Precision
Precision studies were performed in a laboratory using the
ReliOn™ Premier BLU Blood Glucose Monitoring System.
Within Run Precision
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Between Run Precision
Control avg.
Control avg.
Control avg.
60
Within ± 15 mg/dL and Within ± 20%
371/371 (100%)
43 mg/dL
71 mg/dL
135 mg/dL
203 mg/dL
343 mg/dL
36 mg/dL
114 mg/dL
341 mg/dL
SD = 1.9 mg/dL CV = 4.3%
SD = 2.1 mg/dL CV = 2.9%
SD = 3.8 mg/dL CV = 2.8%
SD = 5.2 mg/dL CV = 2.6%
SD = 11 mg/dL CV = 3.2%
SD = 1.4 mg/dL CV = 3.8%
SD = 3.4 mg/dL CV = 3.0%
SD = 8.2 mg/dL CV = 2.4%