Check that supply voltage written on back label (1) corresponds to the one available in the installation place.
The autoclave must be connected by a protection switch to a system supplied with earthing according to the
country safety regulations (see picture no. 7).
WARNING
!
Do not cause any foldings to the supply cable and do not leave anything on it.
NOTE VALID FOR ITALY:
The system must be executed according to the regulations CEI 64-4 and a differential switch (3) must be
installed with the following characteristics before the socket for the autoclave (4) :
-
-
Autoclave GIMA H100 automatic
ELECTRIC CONNECTION
1
Picture no. 6
nominal current
differential sensibility
Picture no. 7
- 59 -
9
:
10 A.
:
0,03