Electric Connection - Gima H100 Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Check that supply voltage written on back label (1) corresponds to the one available in the installation place.
The autoclave must be connected by a protection switch to a system supplied with earthing according to the
country safety regulations (see picture no. 7).
WARNING
!
Do not cause any foldings to the supply cable and do not leave anything on it.
NOTE VALID FOR ITALY:
The system must be executed according to the regulations CEI 64-4 and a differential switch (3) must be
installed with the following characteristics before the socket for the autoclave (4) :
-
-
Autoclave GIMA H100 automatic

ELECTRIC CONNECTION

1
Picture no. 6
nominal current
differential sensibility
Picture no. 7
- 59 -
9
:
10 A.
:
0,03